John Fluke Mfg. Co., Inc. MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 2620A Hydra: CAL ADJ IEEE /5700 DATE: 10-Jul-92 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 213 CONFIGURATION: Fluke 5700A ============================================================================= # # Source: # This procedure is based on the calibration procedure in the User's Manual # dated March 1991, Rev 2. # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 IEEE [SRQ OFF] 1.003 DRAW 7 78 7 78 100 30 50 95 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Connect the UUT to an ac power source. 1.006 DISP Turn the UUT on. 1.006 DISP Warm-up time: 30 minutes. 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 70%. 1.006 DISP 1.006 DISP Connect the Fluke Hydra to IEEE-488 port 1 on the PC. 1.007 DISP Verify that the Hydra is in IEEE-488 communication mode: 1.007 DISP Press SHIFT, then LIST to select COMM on the Hydra. 1.007 DISP If necessary, use the up arrow key to scroll until IEEE 1.007 DISP is shown on the main display. Press ENTER 1.007 DISP to show the IEEE-488 address. Make sure it is not zero. 1.007 DISP If necessary, use the up or down arrow key to select 1.007 DISP a different address then press ENTER 1.008 HEAD PERFORMING SELF TEST... (~15s) 1.009 IEEE [T10000][SPL][SRQ ON]*TST?[10][D10000][I][GTL] 1.010 MATH MEM1 = MEM 1.011 JMPZ 1.014 1.012 HEAD SELF TEST FAILED: STATUS = [MEM] 1.013 JMP 1.157 1.014 HEAD EQUIPMENT SETUP 1.015 DISP 1.015 DISP *************************************************** 1.015 DISP 1.015 DISP To reduce noise pickup by test leads, particularly 1.015 DISP during high Ohms verification, use shielded test 1.015 DISP cables between the Fluke 5700A and Fluke Hydra. 1.015 DISP 1.015 DISP *************************************************** 1.016 DISP Make the following connections to the Input Module: 1.016 DISP [32] 5700A OUTPUT HI to CH1 HI INPUT 1.016 DISP [32] 5700A OUTPUT LO to CH1 LO INPUT 1.016 DISP [32] 5700A SENSE HI to CH11 HI INPUT 1.016 DISP [32] 5700A SENSE LO to CH11 LO INPUT 1.017 DISP The Cal Enable button is located on the right side 1.017 DISP of the display and is recessed beneath the calibration 1.017 DISP seal. [DRAW 50,90,F] 1.017 DISP Enable the calibration mode by pressing the 1.017 DISP Cal Enable button for about three (3) seconds. 1.017 DISP Use a small screwdriver or equivalent blunt-tipped 1.017 DISP object. Avoid sharper-tipped objects (such as pencils. 1.017 DISP ("CAL" appears in the display.) 1.018 DISP *************************************************** 1.018 DISP In case a time-out error occurs, verify that 1.018 DISP the Fluke Hydra is in the CAL mode! 1.018 DISP *************************************************** 1.019 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEPS 1-6 1.020 IEEE CAL 1 1.021 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 1 1.022 5700 90.000mV S 2W 1.023 IEEE [D7000]CAL_STEP?[I] 1.024 MATH MEM1 = MEM - 0.09 1.025 JMPZ 1.027 1e-006U 1.026 JMP 1.057 1.027 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 2 1.028 5700 0.9000V S 2W 1.029 IEEE [D7000]CAL_STEP?[I] 1.030 MATH MEM1 = MEM - 0.9 1.031 JMPZ 1.033 1e-006U 1.032 JMP 1.057 1.033 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 3 1.034 5700 290.000mV S 2W 1.035 IEEE [D7000]CAL_STEP?[I] 1.036 MATH MEM1 = MEM - 0.29 1.037 JMPZ 1.039 1e-006U 1.038 JMP 1.057 1.039 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 4 1.040 5700 2.9000V S 2W 1.041 IEEE [D7000]CAL_STEP?[I] 1.042 MATH MEM1 = MEM - 2.9 1.043 JMPZ 1.045 1e-006U 1.044 JMP 1.057 1.045 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 5 1.046 5700 29.000V S 2W 1.047 IEEE [D7000]CAL_STEP?[I] 1.048 MATH MEM1 = MEM - 29 1.049 JMPZ 1.051 1e-006U 1.050 JMP 1.057 1.051 HEAD DC VOLTS CALIBRATION ADJUSTMENT STEP 6 1.052 5700 290.00V S 2W 1.053 IEEE [D7000]CAL_STEP?[I] 1.054 MATH MEM1 = MEM - 290 1.055 JMPZ 1.060 1e-006U 1.056 JMP 1.057 1.057 OPBR [MEM]VDC CALIBRATION ADJUSTMENT FAILED! 1.057 OPBR Do you want to repeat DC Volts calibration adjustment? 1.058 JMPT 1.020 1.059 JMP 1.157 1.060 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEPS 7-1 1.061 5700 * S 1.062 IEEE CAL 2 1.063 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 7 1.064 5700 29.000mV 1kH S 2W 1.065 IEEE [D20000]CAL_STEP?[I] 1.066 MATH MEM1 = MEM - 0.029 1.067 JMPZ 1.069 1e-006U 1.068 JMP 1.099 1.069 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 8 1.070 5700 290.00mV 1kH S 2W 1.071 IEEE [D20000]CAL_STEP?[I] 1.072 MATH MEM1 = MEM - 0.29 1.073 JMPZ 1.075 1e-006U 1.074 JMP 1.099 1.075 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 9 1.076 5700 290.00mV 1kH S 2W 1.077 IEEE [D20000]CAL_STEP?[I] 1.078 MATH MEM1 = MEM - 0.29 1.079 JMPZ 1.081 1e-006U 1.080 JMP 1.099 1.081 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 10 1.082 5700 2.9000V 1kH S 2W 1.083 IEEE [D15000]CAL_STEP?[I] 1.084 MATH MEM1 = MEM - 2.9 1.085 JMPZ 1.087 1e-006U 1.086 JMP 1.099 1.087 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 11 1.088 5700 29.000V 1kH S 2W 1.089 IEEE [D15000]CAL_STEP?[I] 1.090 MATH MEM1 = MEM - 29 1.091 JMPZ 1.093 1e-006U 1.092 JMP 1.099 1.093 HEAD AC VOLTS CALIBRATION ADJUSTMENT STEP 12 1.094 5700 290.00V 1kH S 2W 1.095 IEEE [D17000]CAL_STEP?[I] 1.096 MATH MEM1 = MEM - 290 1.097 JMPZ 1.103 1e-006U 1.098 JMP 1.099 1.099 HEAD [MEM]VAC CALIBRATION ADJUSTMENT FAILED! 1.100 OPBR Do you want to repeat AC Volts calibration adjustment? 1.101 JMPT 1.060 1.102 JMP 1.157 1.103 HEAD OHMS CALIBRATION ADJUSTMENT STEPS 13-17 1.104 5700 * S 1.105 IEEE CAL 3 1.106 HEAD OHMS CALIBRATION ADJUSTMENT STEP 13 1.107 5700 190Z S 4W 1.108 IEEE CAL_REF [MEM1] 1.109 IEEE [D15000]CAL_STEP?[I] 1.110 MATH MEM1 = MEM - MEM1 1.111 JMPZ 1.113 0.01U 1.112 JMP 1.141 1.113 HEAD OHMS CALIBRATION ADJUSTMENT STEP 14 1.114 5700 1.9e3Z S 4W 1.115 IEEE CAL_REF [MEM1] 1.116 IEEE [D15000]CAL_STEP?[I] 1.117 MATH MEM1 = MEM - MEM1 1.118 JMPZ 1.120 0.1U 1.119 JMP 1.141 1.120 HEAD OHMS CALIBRATION ADJUSTMENT STEP 15 1.121 5700 1.9e4Z S 4W 1.122 IEEE CAL_REF [MEM1] 1.123 IEEE [D15000]CAL_STEP?[I] 1.124 MATH MEM1 = MEM - MEM1 1.125 JMPZ 1.127 1U 1.126 JMP 1.141 1.127 HEAD OHMS CALIBRATION ADJUSTMENT STEP 16 1.128 5700 1.9e5Z S 4W 1.129 IEEE CAL_REF [MEM1] 1.130 IEEE [D15000]CAL_STEP?[I] 1.131 MATH MEM1 = MEM - MEM1 1.132 JMPZ 1.134 10U 1.133 JMP 1.141 1.134 HEAD OHMS CALIBRATION ADJUSTMENT STEP 17 1.135 5700 1.9e6Z S 4W 1.136 IEEE CAL_REF [MEM1] 1.137 IEEE [D15000]CAL_STEP?[I] 1.138 MATH MEM1 = MEM - MEM1 1.139 JMPZ 1.145 100U 1.140 JMP 1.141 1.141 HEAD [MEM] OHMS CALIBRATION ADJUSTMENT FAILED! 1.142 OPBR Do you want to repeat ohms calibration adjustment? 1.143 JMPT 1.105 1.144 JMP 1.157 1.145 HEAD FREQUENCY CALIBRATION ADJUSTMENT STEP 18 1.146 5700 * S 1.147 IEEE CAL 4 1.148 5700 1.0e4H 2.9V S 2W 1.149 IEEE [D5000] CAL_STEP?[I] 1.150 MATH MEM1 = MEM - 10000 1.151 JMPZ 1.156 1e-006U 1.152 HEAD [MEM]Hz FREQUENCY CALIBRATION ADJUSTMENT FAILED! 1.153 OPBR Do you want to repeat frequency calibration adjustment? 1.154 JMPT 1.147 1.155 JMP 1.157 1.156 HEAD CALIBRATION ADJUSTMENT COMPLETE 1.157 5700 * S 1.158 IEEE *RST;*OPC?[I][GTL] 1.159 END